- The electrode and RTD sensor should be used in a perpendicular position.
- For calibration and recalibration of the pH meter, always use a freshly prepared buffer solution measured at 250VS.
- When calibrating the pH meter, use a buffer close to the expected value of the sample, it will reduce the amplitude error
- The electrode should be washed between sample measurements with deionized or distilled water.
- To remove excess water, never rub the electrode bulb; Simply dry the tip end of the electrode with a clean tissue paper.
- For a fast response, stir the electrodes in the sample or buffer.
- Reactivate or change the electrode with a new one, if the variation in pH 4.00 or pH 9.20 before calibration is greater than 0.3 pH.